OCMA-550 Oil Content Analyser
Description
The OCMA-550 oil content analyser is designed for fast, accurate, and user-friendly oil measurement in both laboratory and field environments. This versatile instrument provides precise analysis of residual oil on components and accurately determines the concentration of oil on solids such as soil, textiles, and machined parts.
Engineered for efficiency, the OCMA-550 oil content analyser delivers results in just 30 to 60 seconds. It uses a highly stable S-316 solvent for extraction, combined with the NDIR (non-dispersive infrared) method to ensure reliable and consistent data. Whether you’re conducting quality checks or performing environmental assessments, this analyser is built to meet a wide range of testing needs.
The quartz sample cell is a key feature that helps keep the instrument clean and reduces the risk of contamination or solvent spills. For enhanced usability, analysis results can be displayed in multiple units—mg/L, mg/kg, or Abs—depending on your specific application.
The OCMA-550 oil content analyser also features a bright, backlit color LCD, making it easy to view measurement results even in low-light environments. Its intuitive interface ensures that both new and experienced users can operate the device with confidence and minimal training.
Whether you’re monitoring oil levels in environmental samples or assessing cleanliness in manufacturing processes, the OCMA-550 oil content analyser offers a reliable, rapid, and simple solution for accurate oil detection.
VIEW OUR ACCESSORIES HERE FOR THE OCMA RANGE
Measurement principle: | Solvent extraction, NDIR analysis (Infrared spectroscopy) |
Measurement range: | 0mg/l ~ 200mg/l dynamic ranging |
Resolution | 0 to 99.9 mg/L; 0.1 mg/L 100 to 200 mg/L; 1 mg/L |
Repeatability | 0 to 9.9 mg/L; ± 0.4 mg/L ± 1 digit 10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit 100 to 200 mg/L; ± 4 mg/L ± 1 digit *For standard liquids |
Display Method | 3.5 inches, 320×240 dots Backlight colour graphic LCD |
Calibration Method | Zero, Span Calibration |
Extraction Solvent | S-316 solvent |
Amount of test sample required | Sample water : Solvent = 2 : 1 |
Amount of extraction solvent required | Approx 6.5ml |
Extraction Method | Extracted manually outside the instrument. |
Ambient OperatingTemperature | 0 to 40 ℃ (non condensation) |
Power supply | AC 100V to 240V ±10%, 50/60 Hz |
Power consumption | AC 240V: Approx.60VA |
External Dimensions | 253(W)×293(D)×195(H) mm |
Mass | Approx. 5 kg |
External Output | Output to an USB memory stick |
Functions | 300-item data memory Stabilised measurement value display Self error determination Clock |