Conductivity Meter with Salinity, TDS and Temp – LAQUA DS-72A-S
$2,570.00 ex GST
For laboratory consumables or equipment, head over to our alternative website, Australian Scientific, where you can conveniently purchase these items alongside water quality products.
$1,365.00 ex GST
Whether you’re performing routine lab work or conducting complex research, the Conductivity Bench Meter delivers the accuracy, reliability, and user-friendly features you need. Built for professionals who demand consistent performance, this meter combines cutting-edge technology with intuitive operation, making it ideal for educational, industrial, and environmental testing settings. From high-resolution readings to smart data management, every feature is tailored to streamline your workflow and enhance confidence in your results.
User-Selectable Resolution
Customize your measurements with resolution settings up to 3 digits for pinpoint accuracy.
Calibration Reminder
Never miss a calibration again! Set the calibration alarm up to 90 days in advance to keep your readings reliable.
Electrode Health Indicator
Quickly assess electrode condition using real-time feedback based on calibration performance.
Stable Measurement Indicator
Know exactly when your readings have settled with stability indicators in both measurement and calibration modes.
Ample Internal Memory
Store up to 2000 sets of data—each tagged with a time and date—for hassle-free tracking and traceability.
Automatic Data Logging
Capture data automatically at custom intervals without lifting a finger. Just set it and forget it.
Secure Setup Access
Keep your settings safe with a 4-digit password that prevents unauthorized adjustments.
Seamless Software Connectivity
Transfer data effortlessly to your PC via USB. Export to CSV, Excel, or PDF using the free DAS20 software.
On-Demand Data Printing
Instantly print results complete with date and time for full GLP/GMP compliance.
Easy Software Updates
Keep your meter up to date with the latest software—download and install upgrades with ease.
Versatile Connectivity
Includes RS-232 and USB ports for quick connection to computers and printers. Enhance your workflow with free downloadable software.
Real-Time Clock & Date Display
Stay GLP-compliant with time and date always visible at the top of the screen.
5” Backlit Digital Display
Bright, easy-to-read screen that shows all critical data and indicators in one view.
Customizable Auto Shut-Off
Save power with adjustable idle shut-off, from a few minutes to 30 minutes.
Flexible Electrode Stand
The built-in, rotatable stand holds up to two electrodes and adjusts for height—perfect for hands-free operation.
3-Year Warranty
Enjoy peace of mind with a full three-year warranty that guarantees durability and dependability. 6 months warrant on electrodes
Need calibration solution for your conductivity bench meter? CLICK HERE to view our range
Model | LAQUA-EC2000 Conductivity/Resistivity/TDS/Salinity/Temp (°C/°F) Bench Meter(°C/°F) Bench Meter |
Conductivity Range | 0.000 to 1.999 µS/cm (k = 0.1) 2.00 to 19.99 µS/cm (k = 0.1, 1) 20.0 to 199.9 µS/cm (k = 0.1, 1, 10) 200 to 1999 µS/cm (k = 0.1, 1, 10) 2.00 to 19.99 mS/cm (k = 0.1, 1, 10) 20.0 to 199.9 mS/cm (k = 1, 10) 0.200 to 2.000 S/cm (k = 10) |
Units | Auto ranging S/cm, S/m (µS ↔ mS ↔ S) |
Resolution | 0.05% full scale |
Accuracy | ± 0.6% full scale; ± 1.5% full scale > 18.0 mS/cm |
Reference Temperature | 15.0 to 30.0 °C (adjustable) |
Temperature Coefficient | 0.00 to 10.00 % per °C (adjustable) |
Cell Constants | 0.070 to 13.00 (adjustable) |
Calibration Points | Up to 4 (Auto) / Up to 5 (Manual) |
Resistivity Range | 0.001 MΩ・cm to 20.0 MΩ・cm |
Resolution | 0.5% full scale |
Accuracy | ± 0.6% full scale; ± 1.5% full scale > 1.80 MΩ・cm |
Total Dissolved Solids (TDS) Range | 0.01 to 9.99 mg/L (ppm) 10.0 to 99.9 mg/L (ppm) 100 to 999 mg/L (ppm) 1.00 to 9.99 g/L (ppt) 10.0 to 100 g/L (ppt) |
Resolution | 0.01, 0.1, 1 mg/L ↔ g/L (ppm ↔ ppt) |
Accuracy | ± 0.1% full scale |
TDS Curves | EN27888, 442, NaCl, Linear (0.40 to 1.00) |
Salinity Range | 0.0 to 100.0 ppt / 0.00 to 10.00 % |
Resolution | 0.1 ppt / 0.01% |
Accuracy | ± 0.2% full scale |
Salinity Curves | NaCl / Seawater |
Calibration Option | Yes (Manual; 1 point) |
Temperature Rangee | -30.0 to 130.0 °C / -22.0 to 266.0 °F |
Resolution | 0.1 °C / °F |
Accuracy | ± 0.5 °C / ± 0.9 °F |
Calibration Option | Yes (± 10.0 °C / ± 18.0 °F range in 0.1 °C increments) |
Memory | 2000 |
Auto Data Log | Yes |
Real-time Clock | Yes |
Date & Time Stamp | Yes |
Measurement Modes | Auto Stable / Auto Hold / Real Time |
Auto Shut-Off | Yes (Programmable: up to 30 mins.) |
Electrode Status | On screen display |
Diagnostics | Yes |
Password Setting | Yes |
Software Upgrade | Yes |
PC / Printer Communication | Phono jack (USB / RS232C) |
Meter Inputs | BNC, phono, DC sockets |
Display | 5″ Custom LCD with backlight and 320 segments |
Power Requirement | AC adaptor 100 – 240V, 50 – 60Hz |
Dimensions & Weight | 155(L) x 150(W) x 67(H) mm, 765g |
CLICK HERE FOR VIEWING AND PRINTING pH CALIBRATION AND DATA
CLICK HERE FOR CONDUCTIVITY CALIBRATION
CLICK HERE FOR VIEWING AND PRINTING CONDUCTIVITY CALIBRATION DATA
CLICK HERE FOR ORP CALIBRATION
CLICK HERE FOR SALINITY CALIBRATION
CLICK HERE FOR VIEWING AND PRINTING SALINITY CALIBRATION DATA